Titel: Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population
VerfasserIn: Langenbucher, Achim
Szentmáry, Nóra
Cayless, Alan
Al Karam, Muntadher
Seiler, Theo G.
Wendelstein, Jascha
Sprache: Englisch
Titel: PloS One
Bandnummer: 20
Heft: 7
Verlag/Plattform: Plos
Erscheinungsjahr: 2025
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: Purpose The purpose of this study was to investigate the repeatability of optical coherence tomography in measuring radius of curvature (ROC) and asphericity (Q) of corneal epithelium, stroma and endothelium in a large patient cohort. Methods In this retrospective non-randomized cross sectional single-center study we eval uated a dataset containing 600 anterior segment optical coherence tomography measurements from 200 eyes (3 repeat measurements each) taken prior to cataract surgery. Measurements were exported containing height map data for the epithelium, stroma and endothelium surface. Model surfaces (floating sphere in a 3 (BFS3) and 6 mm zone (BFS6) and conic surface in a 3 (Conoid3) and 6 mm zone (Conoid6)) were fitted using nonlinear iterative optimization minimizing the height difference between measurement and model. The mean (MEAN) and standard deviations (SD) for each sequence of measurements were derived and analyzed. Results MEAN ROC for epithelium was 7.68/7.72/7.67/7.67 mm with BFS3/BFS6/Conoid3/ Conoid6 and Q was −0.16/-0.19 with Conoid3/Conoid6. Corresponding values for stroma were 7.62/7.64/7.55/7.58 mm and −0.38/-0.24, and for endothelium they were 6.43/6.49/6.40/6.37 mm and −0.29/-0.33. Variations of the 3 repeat measurement ranged between 7–19/18–31/21–32 micron for epithelium/stroma/endothelium, and variation in Q was 0.14/0.20/0.14 with Conoid3 and 0.03/0.05/0.06 with Conoid6. Conclusions Fitting floating sphere or conoid model surfaces to the map data seems to be a very robust and reliable method for extracting ROC and asphericity from all data within a region of interest. For extraction of asphericity, the 6 mm zone should be used due to fewer variations.
DOI der Erstveröffentlichung: 10.1371/journal.pone.0327863
URL der Erstveröffentlichung: https://doi.org/10.1371/journal.pone.0327863
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-482722
hdl:20.500.11880/42216
http://dx.doi.org/10.22028/D291-48272
ISSN: 1932-6203
Datum des Eintrags: 16-Jul-2026
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Augenheilkunde
Professur: M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher
M - Prof. Dr. med. Nóra Szentmáry
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
journal.pone.0327863.pdf1,07 MBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons