Please use this identifier to cite or link to this item:
doi:10.22028/D291-44633
Title: | Performance of a simplified strategy for formula constant optimisation in intraocular lens power calculation |
Author(s): | Langenbucher, Achim Wendelstein, Jascha Szentmáry, Nóra Cayless, Alan Hoffmann, Peter Debellmaniere, Guillaume Gatinel, Damien |
Language: | English |
Title: | Acta Ophthalmologica |
Volume: | 103 (2025) |
Issue: | 1 |
Pages: | e10-e18 |
Publisher/Platform: | Wiley |
Year of Publication: | 2024 |
Free key words: | constant optimisation formula constant formula constant prediction IOL power formula paraxial calculation |
DDC notations: | 610 Medicine and health |
Publikation type: | Journal Article |
Abstract: | Purpose: To investigate the performance of a simple prediction scheme for the formula constants optimised for a mean refractive prediction error. Methods: Analysis based on a dataset of 888 eyes before and after cataract surgery with IOL implantation (Hoya Vivinex). IOLMaster 700 biometric data, power of the implanted lens and postoperative spherical equivalent refraction were used to calculate the optimised constants (.)opt for SRKT, HofferQ, Holladay and Haigis formula with an iterative nonlinear optimisation. For detuning start values by ±1.5 from (.)opt, the predicted formula constants (.)pred were calculated and compared with (.)opt. Formula performance metrics mean (MPE), median (MEDPE), mean absolute (MAPE), median absolute (MEDAPE), root mean squared (RMSPE) and standard deviation (SDPE) of the formula prediction error were analysed for (.)opt and (.)pred. Results: (.)pred – (.)opt showed a 2nd order parabolic behaviour with maximal deviations up to 0.09 at the tails of detuning and a minimal deviation up to −0.01 for all formulae. The performance curves of different metrics of PE as functions of detuning variations show that the formula constants for zeroing MPE and MEDPE yield almost identical formula constants, optimisation for MAPE, MEDAPE and RMSPE yielded formula constants very close to (.)opt, and optimisation for SDPE could result in formula constants up to 0.5 off (.)opt which is unacceptable for clinical use. Conclusion: This simple prediction scheme for formula constant optimisation for zero mean refraction error performs excellently in our monocentric dataset, even for larger deviations of the start value from (.)opt. Further studies with multicentric data and larger sample sizes are required to investigate the performance in a clinical setting further. |
DOI of the first publication: | 10.1111/aos.16692 |
URL of the first publication: | https://doi.org/10.1111/aos.16692 |
Link to this record: | urn:nbn:de:bsz:291--ds-446334 hdl:20.500.11880/39784 http://dx.doi.org/10.22028/D291-44633 |
ISSN: | 1755-3768 1755-375X |
Date of registration: | 13-Mar-2025 |
Description of the related object: | Supporting Information |
Related object: | https://onlinelibrary.wiley.com/action/downloadSupplement?doi=10.1111%2Faos.16692&file=aos16692-sup-0001-DataS1.xlsx |
Faculty: | M - Medizinische Fakultät |
Department: | M - Augenheilkunde |
Professorship: | M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher M - Prof. Dr. med. Nóra Szentmáry |
Collections: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Files for this record:
File | Description | Size | Format | |
---|---|---|---|---|
Acta Ophthalmologica - 2024 - Langenbucher - Performance of a simplified strategy for formula constant optimisation in.pdf | 783,84 kB | Adobe PDF | View/Open |
This item is licensed under a Creative Commons License