Titel: Comparison of reversed-phase, hydrophilic interaction, and porous graphitic carbon chromatography columns for an untargeted toxicometabolomics study in pooled human liver microsomes, rat urine, and rat plasma
VerfasserIn: Hemmer, Selina
Manier, Sascha K.
Wagmann, Lea
Meyer, Markus R.
Sprache: Englisch
Titel: Metabolomics
Bandnummer: 20
Heft: 3
Verlag/Plattform: Springer Nature
Erscheinungsjahr: 2024
Freie Schlagwörter: Untargeted metabolomics
LC-HRMS
Reversed-phase columns
Hydrophilic interaction chromatography columns
Quality assurance
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: Introduction Untargeted metabolomics studies are expected to cover a wide range of compound classes with high chemical diversity and complexity. Thus, optimizing (pre-)analytical parameters such as the analytical liquid chromatography (LC) column is crucial and the selection of the column depends primarily on the study purpose. Objectives The current investigation aimed to compare six diferent analytical columns. First, by comparing the chromatographic resolution of selected compounds. Second, on the outcome of an untargeted toxicometabolomics study using pooled human liver microsomes (pHLM), rat plasma, and rat urine as matrices. Methods Separation and analysis were performed using three diferent reversed-phase (Phenyl-Hexyl, BEH C18, and Gold C18), two hydrophilic interaction chromatography (HILIC) (ammonium-sulfonic acid and sulfobetaine), and one porous graphitic carbon (PGC) columns coupled to high-resolution mass spectrometry (HRMS). Their impact was evaluated based on the column performance and the size of feature count, amongst others. Results All three reversed-phase columns showed a similar performance, whereas the PGC column was superior to both HILIC columns at least for polar compounds. Comparing the size of feature count across all datasets, most features were detected using the Phenyl-Hexyl or sulfobetaine column. Considering the matrices, most signifcant features were detected in urine and pHLM after using the sulfobetaine and in plasma after using the ammonium-sulfonic acid column. Conclusion The results underline that the outcome of this untargeted toxicometabolomic study LC-HRMS metabolomic study was highly infuenced by the analytical column, with the Phenyl-Hexyl or sulfobetaine column being the most suitable. However, column selection may also depend on the investigated compounds as well as on the investigated matrix.
DOI der Erstveröffentlichung: 10.1007/s11306-024-02115-0
URL der Erstveröffentlichung: https://link.springer.com/article/10.1007/s11306-024-02115-0
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-432872
hdl:20.500.11880/38825
http://dx.doi.org/10.22028/D291-43287
ISSN: 1573-3890
Datum des Eintrags: 28-Okt-2024
Bezeichnung des in Beziehung stehenden Objekts: Supplementary Information
In Beziehung stehendes Objekt: https://static-content.springer.com/esm/art%3A10.1007%2Fs11306-024-02115-0/MediaObjects/11306_2024_2115_MOESM1_ESM.xlsx
https://static-content.springer.com/esm/art%3A10.1007%2Fs11306-024-02115-0/MediaObjects/11306_2024_2115_MOESM2_ESM.docx
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Experimentelle und Klinische Pharmakologie und Toxikologie
Professur: M - Prof. Dr. Markus Meyer
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
s11306-024-02115-0.pdf1,55 MBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons