Titel: | Investigation of a Camera-Based Contactless Pulse Oximeter with Time-Division Multiplex Illumination Applied on Piglets for Neonatological Applications |
VerfasserIn: | Thull, René Goedicke-Fritz, Sybelle Schmiech, Daniel Marnach, Aly Müller, Simon Körbel, Christina Laschke, Matthias W. Tutdibi, Erol Nourkami-Tutdibi, Nasenien Kaiser, Elisabeth Weber, Regine Zemlin, Michael Diewald, Andreas R. |
Sprache: | Englisch |
Titel: | Biosensors |
Bandnummer: | 14 |
Heft: | 9 |
Verlag/Plattform: | MDPI |
Erscheinungsjahr: | 2024 |
Freie Schlagwörter: | optical sensors nonlinear dynamical systems image sensors biomedical signal processing biomedical monitoring in vivo neonatology |
DDC-Sachgruppe: | 610 Medizin, Gesundheit |
Dokumenttyp: | Journalartikel / Zeitschriftenartikel |
Abstract: | (1) Objective: This study aims to lay a foundation for noncontact intensive care monitoring of premature babies. (2) Methods: Arterial oxygen saturation and heart rate were measured using a monochrome camera and time-division multiplex controlled lighting at three different wavelengths (660 nm, 810 nm and 940 nm) on a piglet model. (3) Results: Using this camera system and our newly designed algorithm for further analysis, the detection of a heartbeat and the calculation of oxygen saturation were evaluated. In motionless individuals, heartbeat and respiration were separated clearly during light breathing and with only minor intervention. In this case, the mean difference between noncontact and contact saturation measurements was 0.7% (RMSE = 3.8%, MAE = 2.93%). (4) Conclusions: The new sensor was proven effective under ideal animal experimental conditions. The results allow a systematic improvement for the further development of contactless vital sign monitoring systems. The results presented here are a major step towards the development of an incubator with noncontact sensor systems for use in the neonatal intensive care unit. |
DOI der Erstveröffentlichung: | 10.3390/bios14090437 |
URL der Erstveröffentlichung: | https://doi.org/10.3390/bios14090437 |
Link zu diesem Datensatz: | urn:nbn:de:bsz:291--ds-430036 hdl:20.500.11880/38691 http://dx.doi.org/10.22028/D291-43003 |
ISSN: | 2079-6374 |
Datum des Eintrags: | 9-Okt-2024 |
Fakultät: | M - Medizinische Fakultät |
Fachrichtung: | M - Chirurgie M - Pädiatrie |
Professur: | M - Prof. Dr. Michael D. Menger M - Prof. Dr. Michael Zemlin |