Titel: | Effect of different optimization parameters in single isocenter multiple brain metastases radiosurgery |
VerfasserIn: | Altergot, Angelika Ohlmann, Carsten Nüsken, Frank Palm, Jan Hecht, Markus Dzierma, Yvonne |
Sprache: | Englisch |
Titel: | Strahlentherapie und Onkologie |
Bandnummer: | 200 |
Heft: | 9 |
Seiten: | 815-826 |
Verlag/Plattform: | Springer Nature |
Erscheinungsjahr: | 2024 |
Freie Schlagwörter: | Elements Multiple Brain Mets Brain neoplasms Flat vs FFF beams Gradient index Plan complexity |
DDC-Sachgruppe: | 610 Medizin, Gesundheit |
Dokumenttyp: | Journalartikel / Zeitschriftenartikel |
Abstract: | Purpose Automated treatment planning for multiple brain metastases differs from traditional planning approaches. It is therefore helpful to understand which parameters for optimization are available and how they affect the plan quality. This study aims to provide a reference for designing multi-metastases treatment plans and to define quality endpoints for benchmarking the technique from a scientific perspective. Methods In all, 20 patients with a total of 183 lesions were retrospectively planned according to four optimization scenarios. Plan quality was evaluated using common plan quality parameters such as conformity index, gradient index and dose to normal tissue. Therefore, different scenarios with combinations of optimization parameters were evaluated, while taking into account dependence on the number of treated lesions as well as influence of different beams. Results Different scenarios resulted in minor differences in plan quality. With increasing number of lesions, the number of monitor units increased, so did the dose to healthy tissue and the number of interlesional dose bridging in adjacent metastases. Highly modulated cases resulted in 4–10% higher V10% compared to less complex cases, while monitor units did not increase. Changing the energy to a flattening filter free (FFF) beam resulted in lower local V12Gy (whole brain-PTV) and even though the number of monitor units increased by 13–15%, on average 46% shorter treatment times were achieved. Conclusion Although no clinically relevant differences in parameters where found, we identified some variation in the dose distributions of the different scenarios. Less complex scenarios generated visually more dose overlap; therefore, a more complex scenario may be preferred although differences in the quality metrics appear minor. |
DOI der Erstveröffentlichung: | 10.1007/s00066-024-02249-z |
URL der Erstveröffentlichung: | https://link.springer.com/article/10.1007/s00066-024-02249-z |
Link zu diesem Datensatz: | urn:nbn:de:bsz:291--ds-427435 hdl:20.500.11880/38324 http://dx.doi.org/10.22028/D291-42743 |
ISSN: | 1439-099X 0179-7158 |
Datum des Eintrags: | 2-Sep-2024 |
Fakultät: | M - Medizinische Fakultät |
Fachrichtung: | M - Radiologie |
Professur: | M - Prof. Dr. Markus Hecht |
Sammlung: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Dateien zu diesem Datensatz:
Datei | Beschreibung | Größe | Format | |
---|---|---|---|---|
s00066-024-02249-z.pdf | 3,79 MB | Adobe PDF | Öffnen/Anzeigen |
Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons