Please use this identifier to cite or link to this item: doi:10.22028/D291-39004
Title: Implant-Supported Overdentures: Current Status and Preclinical Testing of a Novel Attachment System
Author(s): Wendler, Frank
Diehl, Lisa
Shayanfard, Pejman
Karl, Matthias
Language: English
Title: Journal of Clinical Medicine
Volume: 12
Issue: 3
Publisher/Platform: MDPI
Year of Publication: 2023
Free key words: implant-supported overdenture
patient satisfaction
retention
strain development
NiTi
wear
complications
superelasticity
DDC notations: 610 Medicine and health
Publikation type: Journal Article
Abstract: Numerous attachment systems exist for implant-supported overdentures, with each having specific limitations in terms of retention, cost, wear, maintenance and cleanability. A retrospective analysis of patients restored with implant-supported overdentures using bars, telescopic crowns and Locator-type attachments was performed and the patients were interviewed. An in vitro strain gauge study compared telescopic crowns, Locator-type attachments and a novel flexible attachment system employing a shape memory alloy (NiTi) with respect to peri-implant strain development during insertion, loading and removal of an overdenture. A significantly lower number of attachment-related complications was observed in bars as compared to telescopic crowns (p = 0.00007) and Locator-type attachments (p = 0.00000), respectively. Greater overall patient satisfaction was noted in bar-retained restorations while Locator-type attachments led to lower levels of satisfaction regarding prosthesis retention. In vitro, telescopic crowns caused maximum strain development during prosthesis insertion and loading, while during removal this was observed in Locators with white retentive inserts. NiTi attachments caused significantly lower strain development during insertion as compared to telescopic crowns (p = 0.027). During loading, NiTi attachments caused significantly lower strain development than Locators with blue retentive inserts (p = 0.039). During removal, NiTi attachments caused significantly less strain development as compared to Locators with white retentive inserts (p = 0.027). Positional discrepancies between male and female attachment parts affected the retention and reaction force between both components, which may be minimized by using the novel NiTi attachment system. This may be beneficial in terms of component wear and implant loading.
DOI of the first publication: 10.3390/jcm12031012
URL of the first publication: https://doi.org/10.3390/jcm12031012
Link to this record: urn:nbn:de:bsz:291--ds-390044
hdl:20.500.11880/35179
http://dx.doi.org/10.22028/D291-39004
ISSN: 2077-0383
Date of registration: 13-Feb-2023
Faculty: M - Medizinische Fakultät
Department: M - Zahn-, Mund- und Kieferheilkunde
Professorship: M - Prof. Dr. Matthias Karl
Collections:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Files for this record:
File Description SizeFormat 
jcm-12-01012.pdf2,53 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons