Title: | Highly sensitive benzene detection with metal oxide semiconductor gas sensors – an inter-laboratory comparison |
Author(s): | Sauerwald, Tilman Baur, Tobias Leidinger, Martin Reimringer, Wolfhard Spinelle, Laurent Gerboles, Michel Kok, Gertjan Schütze, Andreas |
Language: | English |
Title: | Journal of sensors and sensor systems : JSSS |
Volume: | 7 |
Issue: | 1 |
Pages: | 235-243 |
Publisher/Platform: | Copernicus Publications |
Year of Publication: | 2018 |
DDC notations: | 621.3 Electrical engineering, electronics |
Publikation type: | Journal Article |
Abstract: | For detection of benzene, a gas sensor system with metal oxide semiconductor (MOS) gas sensors using temperature-cycled operation (TCO) is presented. The system has been tested in two different laboratories at the concentration range from 0.5 up to 10 ppb. The system is equipped with three gas sensors and advanced temperature control and read-out electronics for the extraction of features from the TCO signals. A sensor model is used to describe the sensor response in dependence on the gas concentration. It is based on a linear differential surface reduction (DSR) at a low temperature phase, which is linked to an exponential growth of the sensor conductance. To compensate for cross interference to other gases, the DSR is measured at three different temperatures (200, 250, 300 ∘C) and the calculated features are put into a multilinear regression (partial least square regression – PLSR) for the quantification of benzene at both laboratories. In the tests with the first set-up, benzene was supplied in defined gas profiles in a continuous gas flow with variation of humidity and various interferents, e.g. toluene and carbon monoxide (CO). Depending on the gas background and interferents, the quantification accuracy is between ±0.2 and ±2 ppb. The second gas mixing system is based on a circulation of the carrier gas stream in a closed-loop control for the benzene concentration and other test gases based on continuously available reference measurements for benzene and other organic and inorganic compounds. In this system, a similar accuracy was achieved for low background contaminations and constant humidity; the benzene level could be quantified with an error of less than 0.5 ppb. The transfer of regression models for one laboratory to the other has been tested successfully. |
DOI of the first publication: | 10.5194/jsss-7-235-2018 |
URL of the first publication: | https://jsss.copernicus.org/articles/7/235/2018/ |
Link to this record: | urn:nbn:de:bsz:291--ds-373316 hdl:20.500.11880/33810 http://dx.doi.org/10.22028/D291-37331 |
ISSN: | 2194-878X 2194-8771 |
Date of registration: | 22-Sep-2022 |
Faculty: | NT - Naturwissenschaftlich- Technische Fakultät |
Department: | NT - Systems Engineering |
Professorship: | NT - Prof. Dr. Andreas Schütze |
Collections: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Files for this record:
There are no files associated with this item.
Items in SciDok are protected by copyright, with all rights reserved, unless otherwise indicated.