Titel: Reliability and efficiency of corneal thickness measurements using sterile donor tomography in the eye bank
VerfasserIn: Hamon, Loïc
Quintin, Adrien
Mäurer, Stephanie
Weinstein, Isabel
Langenbucher, Achim
Seitz, Berthold
Daas, Loay
Sprache: Englisch
Titel: Cell and Tissue Banking
Verlag/Plattform: Springer Nature
Erscheinungsjahr: 2021
Freie Schlagwörter: Donor cornea
Eye bank
Reliability
Efficiency
Optical coherence tomography
Organ culture
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: To evaluate the reliability and efficiency of sterile pachymetric measurements of donor corneas based on tomographic data using two different methods: a “manual” and a “(semi-)automated” method. Twenty-five (25) donor corneas (50%) stored in MI and 25 (50%) in MII were imaged 5 times consecutively using an anterior segment OCT (AS-OCT). The central corneal thickness (CCT) was measured both with the manual measurement tool of the AS-OCT (= CCTm) and with a MATLAB self-programmed software allowing (semi-)automated analysis (= CCTa). We analyzed the reliability of CCTm and CCTa using Cronbach´s alpha (α) and Wilcoxon signed-Rank Test. Concerning CCTm, 68 measurements (54.4%) in MI and 46 (36.8%) in MII presented distortions in the imaged 3D-volumes and were discarded. Concerning CCTa, 5 (4%) in MI and 1 (0.8%) in MII were not analyzable. The mean (± SD) CCTm was 1129 ± 6.8 in MI and 820 ± 5.1 µm in MII. The mean CCTa was 1149 ± 2.7 and 811 ± 2.4 µm, respectively. Both methods showed a high reliability with a Cronbach´s α for CCTm of 1.0 (MI/MII) and for CCTa of 0.99 (MI) and 1.0 (MII). Nevertheless, the mean SD of the 5 measurements was significantly higher for CCTm compared to CCTa in MI (p = 0.03), but not in MII (p = 0.92). Sterile donor tomography proves to be highly reliable for assessment of CCT with both methods. However, due to frequent distortions regarding the manual method, the (semi-)automated method is more efficient and should be preferred.
DOI der Erstveröffentlichung: 10.1007/s10561-021-09980-2
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-353781
hdl:20.500.11880/32287
http://dx.doi.org/10.22028/D291-35378
ISSN: 1573-6814
1389-9333
Datum des Eintrags: 2-Feb-2022
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Augenheilkunde
Professur: M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher
M - Prof. Dr. Berthold Seitz
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
Hamon2021_Article_ReliabilityAndEfficiencyOfCorn.pdf1,54 MBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons