Titel: Repeatability of Corneal Astigmatism and Equivalent Power with the MS-39 Tomographer Derived from Model Surface Fitting in a Cataractous Population
VerfasserIn: Langenbucher, Achim
Szentmáry, Nóra
Cayless, Alan
Al Karam, Muntadher
Hoffmann, Peter
Seiler, Theo G.
Wendelstein, Jascha
Sprache: Englisch
Titel: Sensors
Bandnummer: 25
Heft: 19
Verlag/Plattform: MDPI
Erscheinungsjahr: 2025
Freie Schlagwörter: repeatability of astigmatism
repeatability equivalent power
astigmatic surface model fit
three-surface cornea model
epithelial mapping
corneal power vectors
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: We investigated the repeatability of the MS-39 in determining power vector components—the spherical equivalent (SEQ) and astigmatic powers (C0 and C45) and asphericity (Q)—of corneal epithelium, stroma, and endothelium in a large patient cohort. In this retrospective cross-sectional single-centre study, we evaluated a dataset containing 600 MS-39 anterior segment tomography measurements from 200 eyes (three repeat measurements each) taken prior to cataract surgery. The exported measurements included height map data for the epithelium, stroma, and endothelium surface. Model surfaces (spherocylinder (SphCyl), cylindrical conoid (CylConoid), and biconic (Biconic), all in the 3/6 mm zone) were fitted using nonlinear iterative optimisation, minimising the height difference between the measurement and model. The mean (MEAN) and standard deviation (SD) for each sequence of measurements were derived and analysed. In the 3 mm and 6 mm zone, the MEAN SEQ was 53.47/53.56/53.57 and 53.21/53.54/53.54 D for SphCyl/CylConoid/Biconic for the epithelium, −4.47/−4.51/−4.51 and −4.45/−4.50/−4.50 D for the stroma, and −6.23/−6.26/−6.26 and −6.18/−6.29/−6.30 D for the endothelium. With the three surface models and the 3/6 mm zone, the SD for SEQ/C0/C45 was in the range of 0.04 to 0.11/0.05 to 0.13/0.04 to 0.11 D for epithelium; 0.01 to 0.02/0.01 to 0.05/0.01 to 0.06 D for stroma; and 0.01 to 0.02/0.02 to 0.07/0.03 to 0.07 D for endothelium. Fitting floating model surfaces with astigmatism to map data of the corneal epithelium, stroma, and endothelium seems to be a robust and reliable method for extracting equivalent power and astigmatism using all the datapoints within a region of interest.
DOI der Erstveröffentlichung: 10.3390/s25196171
URL der Erstveröffentlichung: https://doi.org/10.3390/s25196171
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-464190
hdl:20.500.11880/40708
http://dx.doi.org/10.22028/D291-46419
ISSN: 1424-8220
Datum des Eintrags: 16-Okt-2025
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Augenheilkunde
Professur: M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher
M - Prof. Dr. med. Nóra Szentmáry
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
sensors-25-06171-v2.pdf3,25 MBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons