Bitte benutzen Sie diese Referenz, um auf diese Ressource zu verweisen: doi:10.22028/D291-45691
Titel: Femtosecond Laser-Assisted Donor and Recipient Preparation for Bowman Layer Transplantation
VerfasserIn: Schlosser, Rosemarie
Rickmann, Annekatrin
Szurman, Peter
Wahl, Silke
Seitz, Berthold
Wakili, Philip
Müller, Lisa Julia
Roberts, Philipp Ken
Boden, Karl Thomas
Sprache: Englisch
Titel: Journal of Clinical Medicine
Bandnummer: 14
Heft: 12
Verlag/Plattform: MDPI
Erscheinungsjahr: 2025
Freie Schlagwörter: Femtosecond Laser (FSL)
Bowman layer transplantation
keratoconus
femtosecond laser Ziemer LDV Z8
low-energy femtosecond laser
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: Objectives: A Ziemer LDV Z8 femtosecond laser (FSL) was used to obtain optimal cutting parameters with precise settings for donor and recipient preparations for Bowman layer transplantation. Methods: Of 48 human research corneas examined, 32 were used for Bowman layer preparation (donor) and 16 for pocket preparation (recipient) using the LDV Z8 FSL. The cutting thickness of the Bowman layer, pocket depth, and corresponding laser settings were varied. The quality of sections was evaluated based on the occurrence of adhesions, bridges, or perforations. Histological specimens were prepared and analyzed. Results: Preparation of the Bowman layer and recipient pocket was possible using all selected settings. The thinner the Bowman layer and the more superficial the pocket preparation, the higher the risk of perforation was. Considering the fact that the Bowman layer was cut as thinly as possible, a Bowman layer thickness of 30 µm showed a 100% success rate. Bowman layers cut at 25 µm had a lower success rate (50%). The pocket depth of 150 µm showed a 100% success rate in the preparation. Histological processing revealed smooth, precisely cut edges of Bowman layers and pockets. Implantation into the pocket was successful in all cases. Conclusions: Both Bowman layer and pocket preparation were technically and surgically feasible using the LDV Z8 FSL, and the prepared Bowman layers were thinner than those reported in previous studies. The optimal Bowman layer thickness was 30 µm, and a resection depth of 150 µm was used to prepare the pockets safely.
DOI der Erstveröffentlichung: 10.3390/jcm14124362
URL der Erstveröffentlichung: https://doi.org/10.3390/jcm14124362
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-456918
hdl:20.500.11880/40190
http://dx.doi.org/10.22028/D291-45691
ISSN: 2077-0383
Datum des Eintrags: 27-Jun-2025
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Augenheilkunde
Professur: M - Prof. Dr. Berthold Seitz
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
jcm-14-04362.pdf1,04 MBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons