Bitte benutzen Sie diese Referenz, um auf diese Ressource zu verweisen: doi:10.22028/D291-44229
Volltext verfügbar? / Dokumentlieferung
Titel: Retrieval of Migrated Volume Coils Using Different Clot Retrievers in a Porcine Model
VerfasserIn: Simgen, Andreas
Tomori, Toshiki
Mühl-Benninghaus, Ruben
Bomberg, Hagen
Yilmaz, Umut
Körner, Heiko
Laschke, Matthias W.
Menger, Michael D.
Reith, Wolfgang
Sprache: Englisch
Titel: Clinical Neuroradiology
Bandnummer: 28 (2018)
Heft: 4
Seiten: 593-600
Verlag/Plattform: Springer Nature
Erscheinungsjahr: 2017
Freie Schlagwörter: PC 400
Trevo ProVue
Solitaire FR
3DSeparator
ERIC
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: Abstract Purpose During endovascular treatment of intracranial aneurysms using coils, migration is a serious complication that increases neurological morbidity. The aim of this experimental study was to assess the effectiveness and complications of retrieving volume coils with different, currently available clot retrievers in a porcine model. Methods Volume coils of three-dimensional (3D) shape and different sizes were placed in the axillary artery of pigs. By means of 4 different clot retrievers (Trevo ProVue; Solitaire FR; 3D-Separator; ERIC) a total of 40 retrieval maneuvers (10 per retriever) were performed by deploying the retrievers within the migrated coils and trying to trap parts of the coils by advancing the microcatheter. Retrieval rates, retrieval duration, retrieval attempts, and complications were assessed. Results Overall coil retrieval was successful in 31 of 40 cases (77.5%). Retrieval rates using the Trevo ProVue (9/10), Solitaire FR (8/10), and the 3D-Separator (9/10) were higher than when using the ERIC (5/10). Duration of retrieval and retrieval attempts were significantly higher using the ERIC (p < 0.05). Complications like inadvertent deployment were only observed in one case using the Solitaire FR. Additional entrapment of the coil-retriever complex at the intermediate catheter was seen in 9 cases (22.5%). There was no case of vasospasm, perforation, or dissection. Conclusions Retrieval of migrated volume coils using newgeneration clot retrievers is a feasible and effective method. Retrieval rates and duration with the Trevo ProVue, Solitaire FR, and 3D-Separator are superior when compared to the ERIC.
DOI der Erstveröffentlichung: 10.1007/s00062-017-0600-1
URL der Erstveröffentlichung: https://link.springer.com/article/10.1007/s00062-017-0600-1
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-442292
hdl:20.500.11880/39537
http://dx.doi.org/10.22028/D291-44229
ISSN: 1869-1447
1869-1439
Datum des Eintrags: 30-Jan-2025
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Chirurgie
M - Radiologie
Professur: M - Prof. Dr. Michael D. Menger
M - Prof. Dr. Wolfgang Reith
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Es gibt keine Dateien zu dieser Ressource.