Titel: | Does standoff material affect acoustic radiation force impulse elastography? A preclinical study of a modified elastography phantom |
VerfasserIn: | Hollerieth, Katharina Gaßmann, Bernhard Wagenpfeil, Stefan Kemmner, Stephan Heemann, Uwe Stock, Konrad Friedrich |
Sprache: | Englisch |
Titel: | Ultrasonography |
Bandnummer: | 37 |
Heft: | 2 |
Seiten: | 140-148 |
Verlag/Plattform: | Korean Society of Ultrasound in Medicine |
Erscheinungsjahr: | 2018 |
Freie Schlagwörter: | Elasticity imaging techniques Ultrasonography Standoff material Muscles Transducers |
DDC-Sachgruppe: | 610 Medizin, Gesundheit |
Dokumenttyp: | Journalartikel / Zeitschriftenartikel |
Abstract: | Purpose: This study was conducted to determine the influence of standoff material on acoustic radiation force impulse (ARFI) measurements in an elasticity phantom by using two different probes. Methods: Using ARFI elastography, 10 observers measured the shear wave velocity (SWV, m/sec) in different lesions of an elasticity phantom with a convex 4C1 probe and a linear 9L4 probe. The experimental setup was expanded by the use of an interposed piece of porcine muscle as standoff material. The probe pressure on the phantom was registered. Results: Faulty ARFI measurements occurred more often when quantifying the hardest lesion (74.0 kPa 4.97 m/sec) by the 9L4 probe with the porcine muscle as a standoff material interposed between the probe and the phantom. The success rate for ARFI measurements in these series was 52.4%, compared with 99.5% in the other series. The SWV values measured with the 9L4 probe were significantly higher (3.33±1.39 m/sec vs. 2.60±0.74 m/sec, P<0.001 in the group without muscle) and were closer to the reference value than those measured with the 4C1 probe (0.25±0.23 m/sec vs. 0.85±1.21 m/sec, P<0.001 in the same group). The SWV values measured when using the muscle as a standoff material were lower than those without the muscle (significant for 9L4, P=0.040). The deviation from the reference value and the variance increased significantly with the 9L4 probe if the muscle was in situ (B=0.27, P=0.004 and B=0.32, P<0.001). In our study, the pressure exerted by the operator had no effect on the SWV values. Conclusion: The presence of porcine muscle acting as a standoff material influenced the occurrence of failed measurements as well as the variance and the accuracy of the measured values. The linear high-frequency probe was particularly affected. |
DOI der Erstveröffentlichung: | 10.14366/usg.17002 |
URL der Erstveröffentlichung: | https://www.e-ultrasonography.org/journal/view.php?doi=10.14366/usg.17002 |
Link zu diesem Datensatz: | urn:nbn:de:bsz:291--ds-406931 hdl:20.500.11880/36578 http://dx.doi.org/10.22028/D291-40693 |
ISSN: | 2288-5919 |
Datum des Eintrags: | 9-Okt-2023 |
Fakultät: | M - Medizinische Fakultät |
Fachrichtung: | M - Medizinische Biometrie, Epidemiologie und medizinische Informatik |
Professur: | M - Prof. Dr. Stefan Wagenpfeil |
Sammlung: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Dateien zu diesem Datensatz:
Datei | Beschreibung | Größe | Format | |
---|---|---|---|---|
usg-17002.pdf | 274,48 kB | Adobe PDF | Öffnen/Anzeigen |
Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons