Titel: Residual volatile anesthetics after workstation preparation and activated charcoal filtration
VerfasserIn: Müller-Wirtz, Lukas M.
Godsch, Christine
Sessler, Daniel I.
Volk, Thomas
Kreuer, Sascha
Hüppe, Tobias
Sprache: Englisch
Titel: Acta Anaesthesiologica Scandinavica
Bandnummer: 64
Heft: 6
Seiten: 759–765
Verlag/Plattform: Wiley
Erscheinungsjahr: 2020
Freie Schlagwörter: activated charcoal filter
anesthesia workstation
desflurane
malignant hyperthermia
MCC-IMS
multicpillary column–ion mobility spectrometry
sevoflurane
trigger-free
vapor-free
washout
DDC-Sachgruppe: 610 Medizin, Gesundheit
Dokumenttyp: Journalartikel / Zeitschriftenartikel
Abstract: Background Volatile anesthetics potentially trigger malignant hyperthermia crises in susceptible patients. We therefore aimed to identify preparation procedures for the Draeger Primus that minimize residual concentrations of desflurane and sevoflurane with and without activated charcoal filtration. Methods A Draeger Primus test workstation was primed with 7% desflurane or 2.5% sevoflurane for 2 hours. Residual anesthetic concentrations were evaluated with five preparation procedures, three fresh gas flow rates, and three distinct applications of activated charcoal filters. Finally, non‐exchangeable and autoclaved parts of the workstation were tested for residual emission of volatile anesthetics. Concentrations were measured by multicapillary column–ion mobility spectrometry with limits of detection/quantification being <1 part per billion (ppb) for desflurane and <2.5 ppb for sevoflurane. Results The best preparation procedure included a flushing period of 10 minutes between removal and replacement of all parts of the ventilator circuit which immediately produced residual concentrations <5 ppm. A fresh gas flow of 10 L/minute reduced residual concentration as effectively as 18 L/minute, whereas flows of 1 or 5 L/minute slowed washout. Use of activated charcoal filters immediately reduced and maintained residual concentrations <5 ppm for up to 24 hours irrespective of previous workstation preparation. The fresh gas hose, circle system, and ventilator diaphragm emitted traces of volatile anesthetics. Conclusion In elective cases, presumably safe concentrations can be obtained by a 10‐minute flush at ≥10 L/minute between removal and replacement all components of the airway circuit. For emergencies, we recommend using an activated charcoal filter.
DOI der Erstveröffentlichung: 10.1111/aas.13571
Link zu diesem Datensatz: urn:nbn:de:bsz:291--ds-332606
hdl:20.500.11880/30609
http://dx.doi.org/10.22028/D291-33260
ISSN: 1399-6576
0001-5172
Datum des Eintrags: 10-Feb-2021
Fakultät: M - Medizinische Fakultät
Fachrichtung: M - Anästhesiologie
Professur: M - Prof. Dr. Thomas Volk
Sammlung:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Dateien zu diesem Datensatz:
Datei Beschreibung GrößeFormat 
aas.13571.pdf641,79 kBAdobe PDFÖffnen/Anzeigen


Diese Ressource wurde unter folgender Copyright-Bestimmung veröffentlicht: Lizenz von Creative Commons Creative Commons